Type |
Type |
| Requirements |
PER 1 |
PEROX |
| Pollution and chemical burdened sample matrix |
Suitable thanks to waterproof diaphragm |
More susceptible as diaphragm not water proof
|
|
Electric interference due to disturbance potential in measuring medium |
Unsusceptible as counter electrode separated from process |
More susceptible as counter electrode is in medium |
| Temperature range |
Up to 50 °C |
Up to 40 °C |
|
Easy handling during installation and maintenance |
Suitable thanks to temperature compensation and measuring transducer integrated into sensor |
Separate temperature sensor and measuring transducer |
|
Response time for H2O2 for fast regulation |
Slow T90 = 6-8 min |
Fast: T90 = 20 sec |
| Fast temperature changes |
Slow due to integrated temperature sensor |
Fast thanks to separate temperature sensor |
|
Long process cycles without presence of H2O2 |
not suitable |
Suitable thanks to pulsated polarisation technology |
|
Measurement range may vary from time to time in various figures or is unclear at the time the order is placed |
Selection of the suitable sensor is necessary |
Suitable as measuring range can be changed manually from the sensor transducer |
| Price of the measuring station |
lower |
higher |