DULCOMETER® D1C Single Channel Controller
Single channel controller with control characteristics P and PID, measuring variables pH, Redox, Chlorine, Bromine, Chlorine Dioxide, Chlorite, Ozone, inductive and conductive conductivity, Fluoride, H2o2, Peracetic Acid, Temperature and Standard Signal.
|
 |
 |
The single channel controller DULCOMETER® D1C is a measuring and control device based on microprocessor technology that grants highest process safety with a comprehensive application spectrum. The controller is configurable and can thus individually be adjusted to your requirements. This is not only valid for its type of control but also for its construction type (wall mounting or panel mounting) and many more parameters. You select what you need for your regulation task and only pay for that.
Advantages
- Universally implementable, as many measuring values controllable: beside pH, Redox, conductivity and Chlorine, also Dissolved Oxygen, Fluoride, Bromine, Chlorine Dioxide, Chlorite, Ozone, Hydrogen Peroxyde and Peracetic Acid.
- Ideal adjustment to process requirements thanks to various stages of extension
- Seamless integration into your process technology: variable installation possibilities for the sensors and large number of control possibilities for metering devices.
- Optimized process flows through special functions such as disturbance feedforward, pH compensation for Chlorine, base load dosing and many limit value functions.
- Safety: limit value monitoring with standardized dosing time restrictions.
- Easy and clear user guidance with text indicator on big graphic display in 16 languages.
- Large, easily readable graphic display for measuring values, correction values and information.
Features
- Limit value control with standardized control variable shutoff
- Disturbance insensitive probe connection in dual wire technology
- 2 signal outputs 0/4 - 20 mA galvanically separated
- Various construction types for wall and panel mounting
- 1 digital input for "break" and "error" measuring water
- Disturbance signal processing as digital or analogue signal
- 2 timers as cleaning contact or for shock dosing
- Additive basic load dosing
- Bidirectional control
- Neutral zone control
- Temperature compensation for pH, conductivity and Chlorine Dioxide process measuring
- pH correction for Chlorine
- Adjustable access code
Technical Data
| Measurand |
Value |
| Accuracy |
0,5 % of measuring range |
| Correction variable |
Temperature over Pt 100 or transmitter 4...20 mA |
| Correction range |
0...100°C |
| Disturbance signals |
Adder/multiplier |
| Control characteristic |
P, PI, PD, PID |
| Control |
Bidirectional control
Dead zone control
Additive basic load
manually adjusted control variable for test purposes |
| Control output |
2 reed contacts (pulse rate output)
2 power relays (pulse length) or
1 output for servomotor with report function
2 analogue outputs (0/4...20 mA) |
| Analogue output |
2 analogue outputs (0/4...20 mA) galvanically separated, max. load 600 Ω (2nd output 400 Ω). Adjustable direction measurement, correction and control variable. |
| Alarm relay |
250V ~3 A, 700 VA, changeover contact |
| Power supply |
24 V AC/DC / 100 V~ / 115 V~ / 200 V~ /230 V~ ±10% |
| Protection class |
Wall mounted: IP 65
Control panel version: IP 54 |
| Ambient temperature |
Wall mounted: -5...40°C (not precipitating)
Control panel version: 0...40 °C (not precipitating) |
| Dimensions in mm (HxWxD) |
Control panel version: 96 x 96 x 140
Wall mounted: 189 x 200 x 76 |
|
Measurand |
Measuring and Control Range |
Resolution |
Sensor input |
| pH |
0 - 14 |
0,01pH |
Input resistance > 1012 W Clamp standard signal 0/4...20 mA |
| Redox |
-1000 - +1000 mV |
1 mV |
Input resistance > 5 x 10 11 W Clamp standard signal 0/4...20 mA |
| Chlorine |
0,005 - 0,500 ppm 0,05 - 2,00 ppm 0,05 - 5,00 ppm 0,05 - 10,00 ppm 0,05 - 20,00 ppm 0,05 - 50,00 ppm 0,05 - 100,00 ppm |
0,001 ppm 0,01 ppm |
Clamp standard signal 0/4...20 mA
|
| Bromine |
0,02 - 2,00 ppm 0,1 - 10,0 ppm |
0,01 ppm |
Clamp standard signal 0/4...20 mA |
| Conductive conductivity |
0 - 20 µS/cm 0 - 200 µS/cm 0 - 2 mS/cm 0 - 200 mS/cm |
0,001 µS/cm |
Clamp conductive conductivity |
| Inductive conductivity |
0 - 200 µS/cm 0 - 2 mS/cm 0 - 200 mS/cm 0 - 2000 mS/cm |
0,01 µS/cm |
Clamp inductive conductivity |
| Chlorine dioxide |
0,005 - 0,500 ppm 0,05 - 2,00 ppm 0,05 - 10,00 ppm 0,05 - 20,00 ppm |
0,001 ppm 0,01 ppm |
Clamp standard signal 0/4...20 mA |
| Chlorite |
0,005 - 0,500 ppm 0,05 - 2,00 ppm |
0,001 ppm 0,01 pmm |
Clamp standard signal 0/4...20 mA |
| Ozone |
0,05 - 2 ppm |
0,01 ppm |
Clamp standard signal 0/4...20 mA |
| Fluoride |
0,05 - 10 mg/l |
0,01 ppm |
Clamp standard signal 0/4...20 mA |
| Dissolved oxygen |
0,1 - 10 mg/l 0,1 - 20 mg/l |
0,001 ppm 0,01 ppm |
Clamp standard signal 0/4...20 mA |
|
Hydrogen peroxide |
1 - 20 ppm 20 - 200 ppm 20 - 2000 ppm 0,01 - 1,00 % |
0,01 ppm
0,01 % |
Clamp standard signal 0/4...20 mA |
| Peracetic acid |
1 - 20 ppm 20 - 200 ppm 20 - 2000 ppm 0,01 - 1,00 % |
0,01 ppm
0,01 % |
Clamp standard signal 0/4...20 mA |
| Temperature |
0 - 100 °C |
0,1 °C |
Clamp Pt 100 / Pt 1000
Clamp standard signal 0/4...20 mA |
| Analogue signal |
0/4...20 mA |
|
Clamp standard signal 0/4...20 mA |
ProMinent offers DULCOMETER® measurement and control devices, DULCOTEST® sensor technology as well as ProMinent metering pumps for the creation of complete control circuits. Measuring, Control, Dosing and Recording all from one hand and optimally inter-adjusted. |
|
|
|
 |