Suggestions

DULCOTEST® Sensors for Total Available Chlorine

Graduated measuring ranges 0.01 – 10 mg/l

Disinfection of swimming pool water with the application-optimised DULCOTEST® sensors for total available chlorine - exact measured values and high process reliability guaranteed.

Product description

Reliable online measurement of total available chlorine – with DULCOTEST® sensors.

Total available chlorine Chlorine bound to (iso)cyanic acid/isocyanurate and the free (effective) chlorine resulting from it. more in the glossary is also known as "stabilised chlorine" and is used to prevent premature decomposition by sunlight in the disinfection of swimming pool water, particularly in outdoor pools. This mainly includes compounds such as derivatives of tri-chloro(iso)cyanuric acid. After it has been metered into the water to be treated, it forms free chlorine and also (iso)cyanuric acid containing chlorine. The sensor measures both compounds, which are designated as total available chlorine Chlorine bound to (iso)cyanic acid/isocyanurate and the free (effective) chlorine resulting from it. more in the glossary.

The innovative diaphragm-covered DULCOTEST® sensor for total available chlorine Chlorine bound to (iso)cyanic acid/isocyanurate and the free (effective) chlorine resulting from it. more in the glossary CGE3 is optimised for the control of the disinfection of swimming pool water with derivatives of chloro(iso)cyanuric acid. The sensor can also be used simultaneously with in-line chlorine electrolysis Chlorine and sodium hydroxide are produced in-situ in electrolysis by passing an electric ... more in the glossary processes without disturbance from the by-products generated.

Your benefits

  • Sensor for total available chlorine, also suitable for the combination of in-line electrolysis with derivatives of chloro(iso)cyanuric acid
  • Can also be used as a sensor for free chlorine without interference caused by the presence of cyanuric acid
  • Precise, real-time amperometric measurement for efficient process control (short response time)
  • Amperometric measuring eliminates interference caused by turbidity or colouration
  • Stable zero point
  • Integrated temperature compensation eliminates interference caused by influence of temperature

Technical Details

Total available chlorine

  • CGE 3-mA
  • pH: 5.5 … 9.5
  • Temperature: 5 … 45 ºC

Sensor for total available and free chlorine CGE 3-mA

Sensor for total available chlorine, such as derivatives of chloro(iso)cyanuric acid, without disturbance when used in swimming pools where disinfection is provided by electrolysis processes. Also suitable for use as a sensor for free chlorine. For operation with controllers with 4-20 mA input

Your benefits

  • Measured variable: total available chlorine, for instance disinfectant with organic chlorine, such as derivatives of chloro(iso)cyanuric acid
  • Measured variable: free chlorine without interference with the presence of cyanuric acid
  • Gold electrode to prevent faults by products from electrolysis processes where the electrodes are immersed directly into the sample water (without diaphragm)
  • Diaphragm-covered sensor (encapsulated) minimises faults caused by changing flow or substances in the water
  • Hydrophilic diaphragm guarantees the permeability of chloro(iso)cyanuric acid derivatives towards the measuring electrodes
  • The special reaction system of the electrolyte allows the total available chlorine to be determined and use at a high pH of up to 9.5

Measured variable

Free chlorine and total available chlorine: Total of organically bound chlorine (e.g. bound to cyanuric acid) and free chlorine

Reference method

DPD1

pH-range

5.5…9.5

Temperature

5…45 °C

Max. pressure

3.0 bar

Flow

DGMa, DLG III: 30...60 l/h
BAMa: 5...100 l/h (depending on design)

Supply voltage

16...24 V DC (2-wire)

Output signal

4-20 mA ≈ measuring range, temperature-compensated, uncalibrated, not electrically isolated

Selectivity

total available chlorine and free chlorine as against combined chlorine (chloramines)

Disinfection process

Disinfectants with organic chlorine, e.g. based on cyanuric acid, chlorine gas, hypochlorite, electrolysis

Process integration

Bypass: open sample water outlet

Sensor fitting

BAMa, DGMa, DLG III

Controllers

D1C, DAC, AEGIS II

Typical applications

Swimming pool water; combined disinfection processes with chloro(iso)cyanuric acid derivatives and electrolysis. Water of a similar quality to potable water with a higher pH of up to 9.5.

Resistance to

Surfactants, cyanuric acid

Measuring principle, technology

Amperometric, 2 electrodes, diaphragm-covered

Measuring range

Order no.

CGE 3-mA-2 ppm

0.02…2.0 mg/l

1047959

CGE 3-mA-10 ppm

0.10…10.0 mg/l

1047975

A mounting kit, order no. 815079, is required for initial fitting of the chlorine sensors in the in-line probe housing DLG III.

Sensor for total available and free chlorine CGE 3-CAN-P

Sensor for total available chlorine, such as derivatives of chloro(iso)cyanuric acid when used in swimming pools. Also suitable for use as a sensor for free chlorine. For use on controllers with CAN-bus connection

Your benefits

  • Measured variable: total available chlorine, for instance disinfectant with organic chlorine, such as derivatives of chloro(iso)cyanuric acid
  • Measured variable: free chlorine without interference with the presence of cyanuric acid
  • Gold electrode to prevent faults by products from electrolysis processes where the electrodes are immersed directly into the sample water (without diaphragm)
  • Diaphragm-covered sensor (encapsulated) minimises faults caused by changing flow or substances in the water
  • Hydrophilic diaphragm guarantees the permeability of chloro(iso)cyanuric acid derivatives towards the measuring electrodes
  • The special reaction system of the electrolyte allows the total available chlorine to be determined and use at a high pH of up to 9.5
  • Operation on the CAN-bus with all the associated benefits

Measured variable

Free chlorine and total available chlorine: Total of organically bound chlorine (e.g. bound to cyanuric acid) and free chlorine

Reference method

DPD1

pH-range

5.5…9.5

Temperature

5…45 °C

Max. pressure

3.0 bar

Flow

DGMa, DLG III: 30...60 l/h
BAMa: 5...100 l/h (depending on design)

Supply voltage

Via CAN interface (11 – 30 V DC)

Output signal

Uncalibrated, temperature compensated, electrically isolated

Selectivity

total available chlorine and free chlorine as against combined chlorine (chloramines)

Disinfection process

Disinfectants with organic chlorine, e.g. based on cyanuric acid, chlorine gas, hypochlorite, electrolysis

Process integration

Bypass: open sample water outlet

Sensor fitting

BAMa, DGMa, DLG III

Controllers

DULCOMARIN® 3, DULCOMARIN® II with hardware before 06.02.2014 from software version 3027 or later, with hardware after 06.02.2014 from software version 3033 or later

Typical applications

Swimming pool water, disinfection processes with chloro(iso)cyanuric acid derivatives and electrolysis. Water of a similar quality to potable water with a higher pH of up to 9.5.

Resistance to

Surfactants, cyanuric acid

Measuring principle, technology

Amperometric, 2 electrodes, diaphragm-covered

Measuring range

Order no.

CGE 3-CAN-P-10 ppm

0.01…10.0 mg/l

1083211

A mounting kit, order no. 815079, is required for initial fitting of the chlorine sensors in the in-line probe housing DLG III.

Downloads for DULCOTEST® Sensors for Total Available Chlorine

More in the Download Centre

0 Results
Filter by document type
Application Example / Reference (0)
Brochure / Flyer (0)
Catalogue (0)
Certificate / Declaration (0)
Data Sheet (0)
Drawing / Dimensional Drawing / 3D (0)
Exploded View / Spare Parts (0)
Material Safety Data Sheet (0)
Operating Instructions (0)
Others (0)
Press Release (0)
Publication / Article (0)
Software (0)
Wiring Diagram (0)

Selected filters:

Matching downloads

Sort by A to Z
Load more matches ...
No more matches available!

Unfortunately, we did not find an appropriate match.

All words written correctly?
Search again with fewer or different search terms?

Please try again later

Service Unavailable

The server is temporarily unable to service your request due to maintenance downtime or capacity problems. Please try again later.

Overview

Reliable online measurement of total available chlorine – with DULCOTEST® sensors.

Total available chlorine Chlorine bound to (iso)cyanic acid/isocyanurate and the free (effective) chlorine resulting from it. more in the glossary is also known as "stabilised chlorine" and is used to prevent premature decomposition by sunlight in the disinfection of swimming pool water, particularly in outdoor pools. This mainly includes compounds such as derivatives of tri-chloro(iso)cyanuric acid. After it has been metered into the water to be treated, it forms free chlorine and also (iso)cyanuric acid containing chlorine. The sensor measures both compounds, which are designated as total available chlorine Chlorine bound to (iso)cyanic acid/isocyanurate and the free (effective) chlorine resulting from it. more in the glossary.

The innovative diaphragm-covered DULCOTEST® sensor for total available chlorine Chlorine bound to (iso)cyanic acid/isocyanurate and the free (effective) chlorine resulting from it. more in the glossary CGE3 is optimised for the control of the disinfection of swimming pool water with derivatives of chloro(iso)cyanuric acid. The sensor can also be used simultaneously with in-line chlorine electrolysis Chlorine and sodium hydroxide are produced in-situ in electrolysis by passing an electric ... more in the glossary processes without disturbance from the by-products generated.

Your benefits

  • Sensor for total available chlorine, also suitable for the combination of in-line electrolysis with derivatives of chloro(iso)cyanuric acid
  • Can also be used as a sensor for free chlorine without interference caused by the presence of cyanuric acid
  • Precise, real-time amperometric measurement for efficient process control (short response time)
  • Amperometric measuring eliminates interference caused by turbidity or colouration
  • Stable zero point
  • Integrated temperature compensation eliminates interference caused by influence of temperature

Application

Technical Data

Downloads